Technology

Specifications

Equine Hyperbaric Chamber

  • Overall Height: 12′
  • Chamber Diameter: 10′
  • Chamber Interior Volume: 825 cubic ft
  • Gross Weight: 24K lbs
  • Nominal Operating Pressure: 30 psig
  • Maximum Allowable Pressure: 33 psig
  • Fabrication Hydrostatic Pressure Tested: 48 psig (>150% nominal operating pressure)

Note: Built in accordance with ASME (American Society of Mechanical Engineers) standards and practices under the strict specifications of Section 8, Division 1

Infrastructure Requirements

  • Electrical
    • One Dedicated 120V, 20 Amp, NEMA 5-20R Receptacle
    • One Switched 120V, 15 AMP NEMA 5-15R Receptacle for Interior Lighting
    • One # 6 AWG Connection to Building Ground System
  • Plumbing
    • Connection for Chamber Drain
    • Connection to Air Supply
    • Chamber Exhaust Penetrations
  • HVAC Maintain Ambient Treatment Room Temperatures at Approximately 68 Degrees

Misc Equipment

  • Air Supply Not Less Than 90 psig. If Not Already Available, Several Standard Models Available at Local Home Improvement or Hardware Stores will Suffice.
  • Foundation. Standard Concrete Preparation As Per Architect or EOT Documentation
  • Lightning Protection. As Per Local Contractor or Ordinance Mandate
  • Standard NFPA Measures. As Per Local Fire Ordinance Where Required

Note: As Per Local and City Ordinances, Site Electrical and Plumbing Connections Must Be Performed By Local Licensed Contractors and are Therefore the Responsibility of the Site Builder