Specifications
Equine Hyperbaric Chamber
-
Overall Height: 12′
-
Chamber Diameter: 10′
-
Chamber Interior Volume: 825 cubic ft
-
Gross Weight: 24K lbs
-
Nominal Operating Pressure: 30 psig
-
Maximum Allowable Pressure: 33 psig
-
Fabrication Hydrostatic Pressure Tested: 48 psig (>150% nominal operating pressure)
Note: Built in accordance with ASME (American Society of Mechanical Engineers) standards and practices under the strict specifications of Section 8, Division 1
Infrastructure Requirements
-
Electrical
-
One Dedicated 120V, 20 Amp, NEMA 5-20R Receptacle
-
One Switched 120V, 15 AMP NEMA 5-15R Receptacle for Interior Lighting
-
One # 6 AWG Connection to Building Ground System
-
Plumbing
-
Connection for Chamber Drain
-
Connection to Air Supply
-
Chamber Exhaust Penetrations
-
HVAC Maintain Ambient Treatment Room Temperatures at Approximately 68 Degrees
Misc Equipment
-
Air Supply Not Less Than 90 psig. If Not Already Available, Several Standard Models Available at Local Home Improvement or Hardware Stores will Suffice.
-
Foundation. Standard Concrete Preparation As Per Architect or EOT Documentation
-
Lightning Protection. As Per Local Contractor or Ordinance Mandate
-
Standard NFPA Measures. As Per Local Fire Ordinance Where Required
Note: As Per Local and City Ordinances, Site Electrical and Plumbing Connections Must Be Performed By Local Licensed Contractors and are Therefore the Responsibility of the Site Builder